Safety Warning
We strongly recommend that you read the general safety information below before using your product for
the first time. If the equipment is not used in the manner specified, then the protection provided may be
impaired. This could result in damage to your computer and/or injury to yourself or others.
Maximum input range
The ADC-100 is designed to measure voltages up to the range -20V to +20V. Any voltages in excess of
±100V may cause permanent damage to the unit.
Mains Voltages
No Pico products are designed for use with mains voltages. To measure mains we recommend the use of
a differential isolating probe specifically designed for such measurements.
Safety Grounding
The ground of every product is connected directly to the ground of your computer via the provided
interconnecting cable. This is done in order to minimise interference. Always use the provided cable to
attach the product to your computer.
As with most oscilloscopes and data loggers, you should take care to avoid connecting the ground input
of the product to anything which may be at some voltage other than ground. If in doubt, use a meter to
check that there is no significant AC or DC voltage. Failure to check may cause damage to the product
and/or computer and could cause injury to yourself or others.
You should assume that the product does not have a protective safety earth. Misconfiguration and/or use
on voltages outside the maximum input range can be hazardous.
Repairs
The unit contains no user serviceable parts: repair or calibration of the unit requires specialised test
equipment and must be performed by Pico Technology Limited or their authorised distributors.